Skip to main content
Fluke Biomedical LANDAUER RaySafe
Survey Meter Study: RaySafe 452 Hybrid Technology versus Ion Chamber Technology
Blog post

Survey Meter Study: RaySafe 452 Hybrid Technology versus Ion Chamber Technology

May 27, 2024
surveymeter-comparison-setup

Introduction

Ensuring radiation safety in medical settings is crucial, particularly for professionals in interventional radiology who are exposed to scattered X-rays. Accurate measurement is essential for safety, traditionally done using ion chamber survey meters. A recent study by Tohoku University researchers evaluates the performance of the RaySafe 452 hybrid survey meter, designed to improve radiation measurement accuracy. Combining a semiconductor detector and a GM tube, it promises better precision and functionality. It was compared with an ion chamber survey meter of another brand to validate its effectiveness.

Link to article about study

Study Findings

  • Reproducibility: The hybrid survey meter has consistent performance, similar to the ion chamber meter across all tube voltages.
  • Energy Dependence: Both technologies are comparable within ±5% across 50-110 kV.
  • Dose Rate Dependence: Differences in readings are within ±2% across various dose rates between the two technologies.
  • Dose Linearity: Nearly equivalent for both technologies, but the hybrid survey meter measures low doses more effectively. It can detect doses as low as 0.013 µSv, surpassing the ion chamber survey meter's detection limit of 0.1 µSv.
  • Angular Dependence: Both technologies have high sensitivity from various angles.

Conclusion

The RaySafe 452 is a valuable tool for measuring scattered radiation, matching and sometimes exceeding the performance of traditional ion chamber survey meters. Its advanced features make it a key device for enhancing radiation safety in medical environments.